• (919) 844 2494
  • Warning Letter: Failure to provide validation (ucm253115)

    Tags:

    Ofni Systems

    Failure to establish and maintain a design history file (DHF) for each type of device as required by 21 CFR 820.30(j). For example, the following documentation was not present: a) Formal documented reviews of the design results, b) The identification, validation, or where appropriate verification of design changes and, c) The results of the design validation, including identification of the design, date, and individuals performing the evaluation.