• (919) 844 2494
  • Warning Letter: Failure to validate (ucm249221)

    Tags:

    Ofni Systems

    You have failed to validate with a high degree of assurance, a process where the results cannot be fully verified by subsequent inspection and test, (21 CFR 820.75(a)). Specifically: You have not validated the preparation of the [redacted] bonding solution which is used to bond the [redacted] subassembly, a component of the Trima ACCEL disposable set; You have not validated the preparation of the [redacted] which is used to bond the [redacted]subassembly and [redacted] subassembly, components of the Spectra disposable sets.

    View the original warning letter.