    Your firm failed to check the accuracy of the input to and output from the computer or related systems of formulas or other records or data and establish the degree and frequency of input/output verifications. For example, the performance qualification of your [redacted] system software (Validation No. 4000-03-PQ-0002) failed to include verification of the expiration date calculations in the [redacted] system. In addition, there is no established degree and frequency of performing the verification. Discrepancy reports have documented that product labeling with incorrect expiration dates have been created and issued for use.