Design validation failed to include testing of production units under actual or simulated use conditions as required by 21 C.F.R. 820.30(g). For example, the MedStar System design history records do not include documentation to demonstrate that the physiological data (e.g. blood pressure and heart rate readings) obtained from the patient is the identical data transmitted from the MedStar Collection Server.