Failure to validate computer software for its intended use according to an established protocol when computers or automated data processing systems are used as part of production or the quality system as required by 21 CFR § 820.70(i). This was a repeat violation from a previous FDA-483 that was issued to your firm. For example: A) Your firm uses off-the-shelf software (HEAT Help Desk) to manage customer support service calls and to maintain customer site configuration information; however, your firm failed to adequately validate this software in order to ensure that it will perform as intended in its chosen application. Specifically. your firm’s validation did not ensure that the details screen was functioning properly as intended. The details screen is used to capture complaint details and complaint follow-up information which would include corrective and preventative actions performed by your firm when service calls are determined to be CAPA issues. B) Off-the-shelf software (Microsoft SharePoint) is being used by your firm to manage your quality system documents for document control and approval. However, your firm has failed to adequately validate this software to ensure that it meets your needs and intended uses. Specifically. at the time of this inspection there were two different versions of your CAPA & Customer Complaint procedure, SOP-200-104; however, no revision history was provided on the SharePoint document history. Your firm has failed to validate the SharePoint software to meet your needs for maintaining document control and versioning.
View the original warning letter.