• (919) 844 2494
  • Warning Letter: Faliure of validation process (ucm170224)

    Tags:

    Ofni Systems

    Your company did not ensure that its design verification and validation process detected design discrepancies with [redacted] the [redacted] printed circuit board (PCB) of the MTS trial stimulators using the [redacted] components that later caused loss of stimulation, and therefore, failure to complete trial implants.