• (919) 844 2494
  • Warning Letter: Failure to validate processes (ucm076566)

    Tags:

    Ofni Systems

    Failure to validate manufacturing processes and approve according to established procedures, where the results cannot be fully verified by subsequent inspections and tests. [21 C.F.R. 820.75(a)] Specifically: Computer Numerical Control (CNC) Machine #C-4 had out of specification results during operational qualification conducted during validation. These out of specification results were not investigated or addressed in the validation report.