• (919) 844 2494
  • Warning Letter: Incomplete documentation of validation (s6741c)

    Tags:

    Ofni Systems

    Failure to adequately establish and maintain procedures for validating the device design, as required by 21 CFR820.30(g). For example:

    a. Device software validation is incomplete. For example, the documentation of the external validation testing for the [redacted] Scanner conducted by a Canadian site in 2005 reveals incomplete sections rating the effectiveness of the scanner.

    b. Discrepancies that were noted at the completion of design validation are not addressed. For example, the final validation report for the [redacted] noted an issue regarding “streak” artifacts at one of the clinical sites; however, there was no further documented evaluation of this artifact to include significance and potential risk prior to the finalization of the design validation and commercial distribution.

    View the original warning letter.


    Recommended Solutions

    Compliance Training >>

    Recommended Solutions

    ExcelSafe >>

    Recommended Solutions

    Ofni Clinical >>

    Recommended Solutions

    Part 11 Toolkit >>

    Recommended Solutions

    Part 11 Advisor >>

    Recommended Solutions

    MedWatch Reporter >>

    Recommended Solutions

    Computer Validation >>

    Recommended Solutions

    Custom Programming >>

    Recommended Solutions

    Validating MS Excel Spreadsheets >>

    Recommended Solutions

    Access Databases >>