The computer software your firm uses to determine metals analysis is deficient. It has no security measures to prevent unauthorized access of the software, no audit trails, and data can be copied or changed at will, with no documentation of the copying or changes. Your procedures do not require the documentation of calculation or entry errors. There is no documentation to indicate that analysts are trained in the software and its applications.